Commissioning
Micropilot M FMR 240 with PROFIBUS PA
60
E Hauser
The basic setup is sufficient for successful commissioning in most applications.
Complex measuring operations necessitate additional functions that the user can use
to customise the Micropilot as necessary to suit his specific requirements. The functions
available to do this are described in detail in the BA 221F.
Comply with the following instructions when configuring the functions in the "
basic
setup
"
(00)
:
•
Select the functions as described on Page 32.
•
Some functions can only be used depending on the parameterisation of the
instrument. For example, the pipe diameter of a stilling well can only be entered if
"
stilling well
" was selected beforehand in the "
tank shape
"
(002)
function.
•
Certain functions (e.g. starting an interference echo mapping (053)) prompt you to
confirm your data entries. Press
O
or
S
to select "
YES
" and press
F
to confirm. The
function is now started.
•
If you do not press a key during a configurable time period (
→
function group
"
display
"
(09)
), an automatic return is made to the home position (measured value
display).
!
Note!
•
The instrument continues to measure while data entry is in progress, i.e. the current
measured values are output via the signal outputs in the normal way.
•
If the envelope curve mode is active on the display, the measured values are updated
in a slower cycle time. Thus, it is advisable to leave the envelope curve mode after the
measuring point has been optimised.
•
If the power supply fails, all preset and parameterised values remain safely stored in
the EEPROM.
"
Caution!
All functions are described in detail, as is the overview of the operating menu itself, in
the manual “
Description of the instrument functions
−
BA 221F
”, which is found on
the enclosed CD-ROM.
!
Note!
The default values of the parameters are typed in
boldface
.