Instrument Settings
(continued)
Table 4-1:
Parameter
Config
Pressure
Standard level psi (kPa)
I/P psi (kPa) @ 4 mA
I/P psi (kPa) @ 20 mA
Keypad input
Serial input
Water vapor
Standard level (wet %/dry
Keypad input %
Measured value
After the calibration is conducted, a cal factor is calculated; this determines the calibration
of the instrument. The cal factor is displayed for a few seconds after a calibration has been
conducted, and you can aslo interrogate it from Mode 4 (
DIAGNOSTIC
mode). Enter the cal
factor into the table below as a record of instrument operation.
Note
Rosemount
™
obtained factory settings under the following conditions:
•
2 m path length
•
Clean conditions
Calibration Data
Table 4-2:
Detector outputs
Factory
Config
D2
D1
E1
E2
Calibration data
Cal factor
Span factor
Output calibration
Set zero
Set span
Operation
Reference Manual
75
Содержание Rosemount CCO 5500
Страница 1: ...Reference Manual 00809 0100 4550 Rev AA June 2018 Rosemount CCO 5500 Carbon Monoxide CO Analyzer ...
Страница 4: ......
Страница 8: ...Contents iv Rosemount CCO 5500 ...
Страница 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Страница 32: ...Install 24 Rosemount CCO 5500 ...
Страница 88: ...Operation 80 Rosemount CCO 5500 ...
Страница 98: ...Maintenance 90 Rosemount CCO 5500 ...
Страница 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Страница 110: ...Spare parts 102 Rosemount CCO 5500 ...
Страница 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Страница 123: ...Return equipment to the factory Reference Manual 115 ...