Select the units for the alarm; these may be different than the units selected for the analog
output.
After you select the source, the instrument requires a level that will trigger the output. Set
the desired level with the arrow keys.
Cal Factor
Note
First, record the original Cal Factor before entering this mode as displayed in Mode 4
→
Calibration
Data. The Cal Factor is lost when the menu option is entered.
During the calibration routine, the instrument calculates a
Cal Factor
which sets the basic
calibration of the instrument. You may change this value from this mode.
Note
Since this value controls the calibration of the instrument, only change if necessary.
Plant Status Input
Use this to determine whether the plant is operating under correct conditions.
There is a choice of three controls for plant status:
Logic Input
,
Serial Input
, and
Multiple
. You
can only use one to control plant status at any one time.
•
Logic Input
If the PS1 and PS2 terminals are linked in the control unit, the logic contact is made,
and the plant status is
OFF
. You may link these terminals manually during a plant
shut down, or you may wire them to a switch/contact outside the unit (e.g., a value
that opens and closes the duct). Press ENTER to select this option when the Plant
Status
→
Serial Input option is displayed.
•
Serial input
Configuration and startup
42
Rosemount CCO 5500
Содержание Rosemount CCO 5500
Страница 1: ...Reference Manual 00809 0100 4550 Rev AA June 2018 Rosemount CCO 5500 Carbon Monoxide CO Analyzer ...
Страница 4: ......
Страница 8: ...Contents iv Rosemount CCO 5500 ...
Страница 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Страница 32: ...Install 24 Rosemount CCO 5500 ...
Страница 88: ...Operation 80 Rosemount CCO 5500 ...
Страница 98: ...Maintenance 90 Rosemount CCO 5500 ...
Страница 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Страница 110: ...Spare parts 102 Rosemount CCO 5500 ...
Страница 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Страница 123: ...Return equipment to the factory Reference Manual 115 ...