6.
Install the air purge unit on the front flange as shown.
7.
Install and tighten the four locking nuts removed in
.
2.5.5
Source and receiver units
Use the following procedure to install the source and receiver units on the air purge units.
1.
Insert a flexible gasket between the air purge unit and the source or receiver unit.
2.
Dowel pins (
) ensure that the source and receiver units and the air purge
units mount in a fixed rotary position. Align the dowel pin and dowel pin hole.
Air Purge and Source Unit Mounting Features
Figure 2-4:
A. Locking nut
B. Window plate
C. Source
D. Alignment dowel pin hole
E. Threaded screw hole
F. Rear face of air purge
3.
Attach the source or receiver to the rear face of the air purge and install the four
screws provided (
).
Install
14
Rosemount CCO 5500