Fault Conditions
(continued)
Table 6-1:
Fault condition
Possible causes
Instrument condition during high opacity condi-
tions which have now cleared.
Low detector levels (D1 < 3,000)
High opacity in duct.
Dirty windows.
Incorrect detector gain adjustment.
Cal factor out of range
Calibration conducted during unstable duct
conditions.
Poor alignment.
Incorrect detector gain adjustment.
Modulation frequency is below 30 Hz or above
45 Hz.
Poor supply voltage.
Faulty chopper motor.
Incorrect setting within the source unit.
Reference signal failure
Chopper motor fails.
Note
If the instrument recognizes a fault condition, it does not update the minutes, hours, and day
averages.
6.2
Troubleshooting tables
The troubleshooting tables provide fault diagnosis, possible causes, and the appropriate
actions if you suspect an instrument fault. Note the symptoms and when the fault has
occurred. Refer to the appropriate tables.
Note
Be sure to use the tables from the top down and pay particular attention when "proceed to next test"
and "proceed to next possible cause" are indicated.
The troubleshooting tables indicate which of the four units failed:
•
Source unit
•
Control unit
•
Receiver unit
•
Power supply unit
You can then return the faulty unit to Rosemount
™
for repair.
Troubleshooting
92
Rosemount CCO 5500
Содержание Rosemount CCO 5500
Страница 1: ...Reference Manual 00809 0100 4550 Rev AA June 2018 Rosemount CCO 5500 Carbon Monoxide CO Analyzer ...
Страница 4: ......
Страница 8: ...Contents iv Rosemount CCO 5500 ...
Страница 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Страница 32: ...Install 24 Rosemount CCO 5500 ...
Страница 88: ...Operation 80 Rosemount CCO 5500 ...
Страница 98: ...Maintenance 90 Rosemount CCO 5500 ...
Страница 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Страница 110: ...Spare parts 102 Rosemount CCO 5500 ...
Страница 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Страница 123: ...Return equipment to the factory Reference Manual 115 ...