3.
Install the source cable in the right front cable port of the control unit enclosure.
Provide adequate free wire length for making connections to the control board
terminals 8 through 12. Tighten the cable gland nut.
4.
Connect the source cable wires to the control board terminals 8 through 12
according to the wiring diagram,
Do not connect the source cable shield wire at the control unit.
5.
Install one end of the power supply to control unit interconnect cable in the center
front cable port of the control unit enclosure. Provide adequate free wire length for
making connections to the control board terminals 1 through 7. Tighten the cable
gland nut.
6.
Connect the cable wires to the control board terminals 1 through 7 according to the
wiring diagram,
.
Do not connect the cable shield wire at the control unit.
7.
Install the opposite end of the power supply to control unit interconnect cable in
one of the right hand cable ports of the power supply unit enclosure. Provide
adequate free wire length for making connections to the power supply board
terminals 1 through 7. Tighten the cable gland nut.
8.
Connect the cable wires to the power supply board terminals 1 through 7 according
to the wiring diagram,
Do not connect the cable shield wire at the power supply unit.
9.
Install the 110/220 Vac power cable in one of the left hand cable ports of the power
supply unit enclosure. Provide adequate free wire length for making connections to
the power supply board terminals: L, N, and E. Tighten the cable gland nut.
10.
Connect the cable wires to the power supply board terminals L, N, and E according
to the wiring diagram,
Do not connect the power cable to the facility power source at this time.
11.
Verify that the power switch is in the correct position.
The voltage position selected must match the voltage supplied to the Rosemount
™
CCO 5500 Analyzer at your facility.
12.
Connect two separate earth ground leads to the ground screws located on the left
hand side of the power supply unit and control unit enclosures.
NOTICE
To maintain proper earth grounding, ensure a positive connection exists between the
enclosures and the earth. The connecting ground wires must be 14 AWG minimum.
Install
Reference Manual
23
Содержание Rosemount CCO 5500
Страница 1: ...Reference Manual 00809 0100 4550 Rev AA June 2018 Rosemount CCO 5500 Carbon Monoxide CO Analyzer ...
Страница 4: ......
Страница 8: ...Contents iv Rosemount CCO 5500 ...
Страница 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Страница 32: ...Install 24 Rosemount CCO 5500 ...
Страница 88: ...Operation 80 Rosemount CCO 5500 ...
Страница 98: ...Maintenance 90 Rosemount CCO 5500 ...
Страница 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Страница 110: ...Spare parts 102 Rosemount CCO 5500 ...
Страница 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Страница 123: ...Return equipment to the factory Reference Manual 115 ...