4.
Unscrew the three brass screws that secure the circuit board and center plate
assembly.
5.
Lift the circuit board and center plate assembly out of the source unit.
6.
De-solder the red and black chopper motor power wires from the circuit board.
7.
Unscrew and remove the three chopper motor screws (
).
Chopper Motor Assembly
Figure 5-6:
A. Chopper motor assembly
B. Chopper motor retaining screws
C. Center plate
8.
Install the new chopper motor and retaining screws removed in
.
9.
Solder the chopper motor power wires to the circuit board. Solder the red wire to
the M+ terminal and the black wire to the M- terminal.
10.
Install the center plate and circuit board assembly. Secure the assembly with the
three brass crews removed in
.
11.
Line up and install the window plate onto the source unit. Fully seat the window
plate flange in the source unit.
12.
Secure the source unit to the air purge using the four screws removed in
.
Tighten the screws evenly.
13.
Power up the Rosemount CCO 5500 Analyzer and check the chopper motor
frequency in Mode
4
(
DIAGNOSTICS
). Adjust using the trim potentiometer as
described in
Maintenance
Reference Manual
85
Содержание Rosemount CCO 5500
Страница 1: ...Reference Manual 00809 0100 4550 Rev AA June 2018 Rosemount CCO 5500 Carbon Monoxide CO Analyzer ...
Страница 4: ......
Страница 8: ...Contents iv Rosemount CCO 5500 ...
Страница 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Страница 32: ...Install 24 Rosemount CCO 5500 ...
Страница 88: ...Operation 80 Rosemount CCO 5500 ...
Страница 98: ...Maintenance 90 Rosemount CCO 5500 ...
Страница 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Страница 110: ...Spare parts 102 Rosemount CCO 5500 ...
Страница 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Страница 123: ...Return equipment to the factory Reference Manual 115 ...