5
Maintenance
5.1
Preventative maintenance
Rosemount designs this equipment to keep maintenance to an absolute minimum.
5.1.1
Cleaning windows
It is important that you keep the optical windows of both the source and receiver
reasonably clean. Keep any mounting tubes free from build-up of dust and fly ash.
Clean the optical window (
) every six months and more frequently for dirty
processes.
Optical Window
Figure 5-1:
A.
Air purge
B.
Source
C.
Window
WARNING!
HIGH TEMPERATURE AND DANGEROUS VAPORS
Take great care when removing the Rosemount
™
CCO 5500 Analyzer from a positive pressure
stack or duct. The source may be very hot, and there may be dangerous vapors present.
Observe all required safety practices.
Procedure
1.
Unbolt and remove the source and receiver from their air purges.
2.
Wipe the windows with a soft dry cloth.
Maintenance
Reference Manual
81