Check Cell Holder
Figure 4-4:
A. Check cell cover
B. Receiver
C. Check cell holder
5.
Insert the check cell into the check cell holder and replace and tighten the screws.
You can insert the cell in either direction. Refer to
Check Cell
Figure 4-5:
A. Gas cell window
Introducing the check cell may cause an initial major disturbance to the instrument
operation.
6.
Wait for the instrument reading to settle (five to ten minutes) and record the gas
measurement with the cell in position.
7.
Remove the check cell and wait for the analyzer to return to zero (another five to ten
minutes).
8.
Replace the cover on the check cell holder and place MODE on the control unit.
The instrument now returns to
OPERATING
mode.
Operation
78
Rosemount CCO 5500