After the isolating valves are installed in the site mounting flanges, connect the purge air
supply and install air purge units according to the following instructions.
2.5.3
Purge air supply
The purpose of the purge air is to keep the windows of the source and the receiver clean.
Always connect the purge air supply to the air purge units before you install the air purge
units on the process duct. Purge air may be supplied by one of the following three
methods.
Negative pressure duct
If the duct operates at a negative pressure under all firing conditions, you may simply leave
the air purge inlets open and allow the negative draft in the duct to draw in ambient air.
You must supply the air purge units for positive pressure ducts with compressed air or
blower air to prevent contamination of the source and receiver units.
Compressed air
You may use compressed air to provide the air flow required. An air supply of 14.7 psi (1
bar) is required, and the consumption is 2.2 cfm (1 liter/second) per purge. Use a fine
control flow regulator and filter.
Blower air
You may use a blower to provide the air to the air purge. Customers may specify their own
blower. The blower should deliver 11 cfm (5 liters/second) per purge against the working
pressure of the duct.
Install
12
Rosemount CCO 5500
Содержание Rosemount CCO 5500
Страница 1: ...Reference Manual 00809 0100 4550 Rev AA June 2018 Rosemount CCO 5500 Carbon Monoxide CO Analyzer ...
Страница 4: ......
Страница 8: ...Contents iv Rosemount CCO 5500 ...
Страница 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Страница 32: ...Install 24 Rosemount CCO 5500 ...
Страница 88: ...Operation 80 Rosemount CCO 5500 ...
Страница 98: ...Maintenance 90 Rosemount CCO 5500 ...
Страница 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Страница 110: ...Spare parts 102 Rosemount CCO 5500 ...
Страница 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Страница 123: ...Return equipment to the factory Reference Manual 115 ...