supplied to the purges until the full installation is complete. If you do any servicing or
rewiring, ensure that the power supply is isolated. During configuration, the system
requires electrical power, compressed air, and open isolating valves.
2.3
Cable requirements
1.
Power supply to signal processor - seven-core, overall screen, multi-stranded, 6/0.2
mm. 0.5 mm
2
.
NOTICE
Although shielded cable is specified for the interconnecting cable, it is not necessary to
ground the cable.
2.
Current loop output - any suitable two-conductor cable, maximum length depends
on keeping output load within the 500 ohm maximum load requirement.
3.
Contact outputs - any two-conductor cable capable of supplying the power to the
warning device/relay, etc. 250 V, 10 A maximum.
4.
A.C. power - any suitable three-conductor power cable capable of transmitting 50
VA.
5.
Serial data link (if required) - twin twisted pair shielded cable.
6.
Analog inputs - any suitable two-conductor cable; Rosemount
™
instruments have an
internal impedance of 240 ohms for these inputs.
2.4
Selecting location
Rosemount
™
designed the equipment for mounting on boiler ducting or stacks open to
the weather. The instrument is fully sealed and requires no further enclosures or
protection. The specific location of the instrument depends on the application and user
requirements. Consider the following when choosing a site.
for a typical system arrangement.
1.
The site must be accessible at both sides of the duct for servicing the source and
receiver.
2.
The site should be as free from extremes of temperature and vibration as possible.
Permissible ambient temperature range is -4 °F to 158 °F (-20 °C to 70 °C).
3.
Flue gas temperatures should not exceed 572 °F (300 °C) at the point of
measurment.
NOTICE
The instrument can achieve temperatures up to 1200 °F (600 °C), but degradation in
instrument accuracy will occur.
4.
There must be an uninterrupted sight path available between the source and
receiver.
Install
8
Rosemount CCO 5500
Содержание Rosemount CCO 5500
Страница 1: ...Reference Manual 00809 0100 4550 Rev AA June 2018 Rosemount CCO 5500 Carbon Monoxide CO Analyzer ...
Страница 4: ......
Страница 8: ...Contents iv Rosemount CCO 5500 ...
Страница 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Страница 32: ...Install 24 Rosemount CCO 5500 ...
Страница 88: ...Operation 80 Rosemount CCO 5500 ...
Страница 98: ...Maintenance 90 Rosemount CCO 5500 ...
Страница 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Страница 110: ...Spare parts 102 Rosemount CCO 5500 ...
Страница 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Страница 123: ...Return equipment to the factory Reference Manual 115 ...