4.11
Check Cell mode
Use this mode with a Rosemount
™
check cell. It verifies calibration and operation of the
analyzer.
For details for this mode's use, refer to
Procedure
1.
Press ENTER on viewing this display if you need to use the
CHECK CELL
mode.
2.
Use the arrow keys to toggle the display to
YES
and press ENTER again.
You have now selected the
CHECK CELL
mode. Note that the display will default back
to Mode
1
if you don't select
YES
within five seconds. Do not insert a the check cell
before accessing this option. Refer to
.
4.12
Shutdown procedure
You do not need to shut down the system unless you are performing maintenance.
1.
In this case, power down the power supply (and thus the control unit and analyzer
heads).
2.
Close the manually operated isolating valves on the analyzer heads.
3.
Proceed with service.
4.
When completed, open the valves and power up the system.
Postrequisites
You may need to calibrate the system depending on the type of maintenance you are
doing. For details, refer to
4.13
Routine checks
4.13.1
Notes for using a Rosemount
™
check cell
The check cell and holder are optional items available from Rosemount. Rosemount has
designed the check cell to verify the reading of Rosemount cross duct analyzers. When
placed within the measurement path, the check cell can generate a known increase in gas
concentration.
Operation
76
Rosemount CCO 5500
Содержание Rosemount CCO 5500
Страница 1: ...Reference Manual 00809 0100 4550 Rev AA June 2018 Rosemount CCO 5500 Carbon Monoxide CO Analyzer ...
Страница 4: ......
Страница 8: ...Contents iv Rosemount CCO 5500 ...
Страница 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Страница 32: ...Install 24 Rosemount CCO 5500 ...
Страница 88: ...Operation 80 Rosemount CCO 5500 ...
Страница 98: ...Maintenance 90 Rosemount CCO 5500 ...
Страница 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Страница 110: ...Spare parts 102 Rosemount CCO 5500 ...
Страница 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Страница 123: ...Return equipment to the factory Reference Manual 115 ...