93
10. Menu System
Serial Interface -
for remote control via PC.
Baud
rate (for COM-Port)- Baud rate of the serial interface. The
baud rate of the opposite terminal must be set to the same
value. Possible values: 4800, 9600, 19200, 38400, 57600,
11520.
Standard setting COM port: Baud: 115200, Data bits: 8, Stop
pulse: 1, Parity: none, Check sum on remote control: off.
Checksum
Checksum on/ off.
Digital outputs / Digital Inputs / Analog outputs / Analog inputs
- (only
for tester with integrated DAS- interface - option E07).
change -
is showing a listing with the outputs or inputs. It is pos-
sible to set individual names for the Out-/input
Fig. 10.41: Setup part 4 (DAS- interface - option E07 only)
Fig. 10.42: Setup part 5
Device-Setup
- for internal parameter. This parameter are set and edit
by the ELABO-service, only. Of this reason the sub menu is password
protected.
Combi-tester G7-1G/-1H/-1T/-1U
Содержание G7-1G
Страница 5: ...5 1 INTRODUCTION free page Combi tester G7 1G 1H 1T 1U...
Страница 38: ...38 4 Operating modes 4 2 Normal Mode Test sequence Combi tester G7 1G 1H 1T 1U...
Страница 39: ...39 4 Operating modes 4 2 1 HV IS test normal mode Combi tester G7 1G 1H 1T 1U...
Страница 40: ...40 4 Operating modes 4 2 2 PE test normal mode Combi tester G7 1G 1H 1T 1U...
Страница 41: ...41 4 Operating modes 4 2 3 Normal mode input step Combi tester G7 1G 1H 1T 1U...
Страница 42: ...42 4 Operating modes 4 2 4 Normal mode output step Combi tester G7 1G 1H 1T 1U...
Страница 43: ...43 4 Operating modes 4 2 4 Normal mode time step Combi tester G7 1G 1H 1T 1U...
Страница 44: ...44 4 Operating modes 4 3 Fast Start Mode test sequence Combi tester G7 1G 1H 1T 1U...
Страница 45: ...45 4 Operating modes 4 3 1 Fast Start Mode PE HV ISO Test Combi tester G7 1G 1H 1T 1U...