33
3. MAKEREADY
3.12.3
STATES PE
Note:
Define the value for the test current and the Rmax limit, as well as
any other parameters using the touch-display or interface (remote control
mode).
State PE READY
The device is ready, i.e. all prerequisites are met to undertake the PE test
(see also table further forward in this document).
State PE ON
Manual start using PE probe (digital input on interface X7). Press
the PE probe onto an electrically conductive part of the test
object.
Automatic start via interface: The PE test is started by the control
computer using an interface command.
Tester ON
PE-test probe
digital Start X11
No
Test current ON
retract
PE-Test probe /
digital signal or
$SHT
Test time finished
Yes
Test current OFF
Start
via RS232
$SST
Set PE-test mode and
PE-test parameter
Test object
faulty
Yes
No
No
yes
No
Yes
Yes
Ready for
operation
Stand by
PE ON
Start via
PE-test probe
Start via
RS232
Digital Start X7
No
Yes
Start via
digital signal
Fig. 3.16: PE start via PE-probe or interface
- no
Combi-tester G7-1G/-1H/-1T/-1U
Содержание G7-1G
Страница 5: ...5 1 INTRODUCTION free page Combi tester G7 1G 1H 1T 1U...
Страница 38: ...38 4 Operating modes 4 2 Normal Mode Test sequence Combi tester G7 1G 1H 1T 1U...
Страница 39: ...39 4 Operating modes 4 2 1 HV IS test normal mode Combi tester G7 1G 1H 1T 1U...
Страница 40: ...40 4 Operating modes 4 2 2 PE test normal mode Combi tester G7 1G 1H 1T 1U...
Страница 41: ...41 4 Operating modes 4 2 3 Normal mode input step Combi tester G7 1G 1H 1T 1U...
Страница 42: ...42 4 Operating modes 4 2 4 Normal mode output step Combi tester G7 1G 1H 1T 1U...
Страница 43: ...43 4 Operating modes 4 2 4 Normal mode time step Combi tester G7 1G 1H 1T 1U...
Страница 44: ...44 4 Operating modes 4 3 Fast Start Mode test sequence Combi tester G7 1G 1H 1T 1U...
Страница 45: ...45 4 Operating modes 4 3 1 Fast Start Mode PE HV ISO Test Combi tester G7 1G 1H 1T 1U...