81
10. Menu System
10.5.2
Automatic test
"test"
- starts the automatic test with the first parameter set in the test plan
The test parameters are displayed in the right column of the status screen
Fig. 10.25: Test plan - start automatic test
Fig. 10.26: Test plan - first parameter set
Start testing:
The device now waits for the manual confirmation of the
start. This confirmation can be given using the green RUN button on the
device or by closing the protective hood on the test cabinet.
In the case of operation via the front panel (local) the green RUN button
(S3/H3) on the device must be pressed twice (2nd start), on this topic also
see chapter 3.12.1. The 2nd-start must be carry out within 30 s. After that
the device interrupts the test with error (confirmation required).
test
Combi-tester G7-1G/-1H/-1T/-1U
HV_1500 AC ....
HV_2000 AC ....
ins/del
ins/del
ins/del
ins/del
test
2HV AC
test plan
first Parameter set
The device is loading the first parameter set ot the test plan (eg. HV_1500
AC).
0.00 [kV]
00.00 [mA]
000.0 [s]
2HV AC
Status: Ready for test!
Start test with RUN!
: HV (AC)
: 2.00 kV
: 0.10 kV/s
: off
: off
: 1.70 kV
: 6.0 s
: off
: 10.0 mA
: 2.00 mA
mode
U
Umin
t
delay
Imax
Imin
1/2
Содержание G7-1G
Страница 5: ...5 1 INTRODUCTION free page Combi tester G7 1G 1H 1T 1U...
Страница 38: ...38 4 Operating modes 4 2 Normal Mode Test sequence Combi tester G7 1G 1H 1T 1U...
Страница 39: ...39 4 Operating modes 4 2 1 HV IS test normal mode Combi tester G7 1G 1H 1T 1U...
Страница 40: ...40 4 Operating modes 4 2 2 PE test normal mode Combi tester G7 1G 1H 1T 1U...
Страница 41: ...41 4 Operating modes 4 2 3 Normal mode input step Combi tester G7 1G 1H 1T 1U...
Страница 42: ...42 4 Operating modes 4 2 4 Normal mode output step Combi tester G7 1G 1H 1T 1U...
Страница 43: ...43 4 Operating modes 4 2 4 Normal mode time step Combi tester G7 1G 1H 1T 1U...
Страница 44: ...44 4 Operating modes 4 3 Fast Start Mode test sequence Combi tester G7 1G 1H 1T 1U...
Страница 45: ...45 4 Operating modes 4 3 1 Fast Start Mode PE HV ISO Test Combi tester G7 1G 1H 1T 1U...