11. Test parameter
Combi-tester G7-1G/-1H/-1T/-1U
11.4 Setting parameters Burn (Option)
BURN with AC voltage, for the localisation of low insulation points inside
the test object. By the over-flash of the voltage, it is possible, to welding
out the insulation fault. The maximum short-circuit current at BURN is:
appr. 250 mA in 2,5 kV range and appr. 150 mA in 5 kV range.
The BURN mode, don't ensures a non destructive testing, since the tripping
current release is blocked. Use the BURN function only for the localisation
an welding out of low insulation points.
test mode ..............
voltage range
test voltage (kV)
run-up (kV/s)
0,5
1,5
low
burn
Burn
Test mode "HV / ISO / PE / Burn / Time"
HV
- HV-test with AC voltage max. 5000 V / max. 200 mA.
ISO
- Insulation-test with DC voltage 1200 V max. 250 MOhm.
PE
-
protective earth conductor test
with AC voltage 6/12 V /
5 ... 32 A / 0 ... 1200 mOhm.
Burn
(option) - HV-test with AC voltage max. 5000 V with the maximum
of test current. The over current tripping is not active.
Time
- Wa i t i n g t i m e s t e p , e g . t i m e t o d i s c h a r g -
ing the test object past a ISO test.
Voltage range "low / high"
low
- test voltage max. 100 .... 2500 V AC
high
- test voltage max. 200 .... 5000 V AC
Test voltage "0.01 ... 5.0 kV AC" -
set the test voltage for the HV-test .
Run up (on/off)
- activates the ramp-up function of the test voltage.
off-
no ramp-up function, the full test voltage is applied.
on
- ramp-up function active. The test voltage is increased linearly
from the start voltage up to the test voltage.
Fig. 11.4a: Test parameter 1 - burn
Содержание G7-1G
Страница 5: ...5 1 INTRODUCTION free page Combi tester G7 1G 1H 1T 1U...
Страница 38: ...38 4 Operating modes 4 2 Normal Mode Test sequence Combi tester G7 1G 1H 1T 1U...
Страница 39: ...39 4 Operating modes 4 2 1 HV IS test normal mode Combi tester G7 1G 1H 1T 1U...
Страница 40: ...40 4 Operating modes 4 2 2 PE test normal mode Combi tester G7 1G 1H 1T 1U...
Страница 41: ...41 4 Operating modes 4 2 3 Normal mode input step Combi tester G7 1G 1H 1T 1U...
Страница 42: ...42 4 Operating modes 4 2 4 Normal mode output step Combi tester G7 1G 1H 1T 1U...
Страница 43: ...43 4 Operating modes 4 2 4 Normal mode time step Combi tester G7 1G 1H 1T 1U...
Страница 44: ...44 4 Operating modes 4 3 Fast Start Mode test sequence Combi tester G7 1G 1H 1T 1U...
Страница 45: ...45 4 Operating modes 4 3 1 Fast Start Mode PE HV ISO Test Combi tester G7 1G 1H 1T 1U...