101
11. Test parameter
Combi-tester G7-1G/-1H/-1T/-1U
Rmax (1 ... 1000 MOhm)
: - largest insulation resistance allowed. If this
value is exceeded, a fault (no contact R > Rmax)) is indicated. This
parameter is used for the contact monitoring. The Rmax is dependent
on the measurement range and can be between 1 and 1000 MOhm.
The ranges are depended on the device type.
Rmin (1 ... dynamic)
: - lowest insulation resistance allowed. If the resist-
ance is less than this value, an insulation fault R<Rmin is indicated. The
Rmin is dependent on the measurement range and can be between 1
and 1000 MOhm. The ranges are depended on the device type.
.
Rmax (MOhm)
Rmin (MOhm)
3.00
45.00
IS_500V
Fig. 11.2d: ISO test parameter 3
Rmax (MOhm)
Rmin (MOhm)
3.00
45.00
IS_500V