103
11. Test parameter
Fig. 11.3b: PE test parameter 2
Combi-tester G7-1G/-1H/-1T/-1U
Tripping value (U, R):
R resistance (0 .... 1.2 Ohm)
nach EN 60601 (VDE 0411).
Tripping limit value
- R resistance (Ohm): - The largest resist-
ance allowed to the protective earth conductor connection. If
this value is exceeded, a PE fault is indicated.
U voltage drop (0 .... 12 V)
acc. EN 60335 (VDE 0700).
Tripping limit value -
voltage drop: - definition of the maximum
voltage drop over the protective earth conductor connection. If
this value is exceeded, a PE fault is indicated.
Start by PE probe
- manual PE testing, start using PE test probe. The
PE test step waits until the PE test probe is pressed against the test
object (internal switch). Only then is the test current applied.