90
10. Menu System
measurement, test voltage not reached, etc.)
fault at: bonding, manual time out, mains inverse polarity,
mains supply not within tolerance.
Buzzer volume
- Loudness of the buzzer in percent.
Passwords -
Activate and change Passwords for the levels 1, 2, 3 (as-
signments of the password steps and predefined passwords - see Chapt.
3.9). The passwords for the layers are individually set able.
Note:
For activating the menu "Settings - Password" the level-3 password
for "Settings" is necessary.
Test- level-1
• Enter the new password for Test (see Chapt. 3.9).
• Accept the new password by using the "save"- button.
The new password is now active.
Parameter - level-2 / Setup - level-3
Enter the passwords for "Parameter" and Setup" in the same way
then "Measure".
Pass words activ
inactive
- no password is active.
active
- passwords active. For starting a new function, it is neces-
sary to enter the password.
Note - Pass words lost:
Request the ELABO service for a "mas-
ter passwords". By using the master password it is possible to
unlock the device.
Device information
- Shows the device number, the software version and
the integrated hardware of the device.
Cabinet impuls
Off
- operation without test cabinet.
On
- for ELABO 93-4B series test cabinet. The device device
sends a pulse to the test cabinet and as a result opens the
lock. The pulse is generated at the end of the test time or after
a „reset“. This is necessary if the device is used in conjunc-
tion with an ELABO 93-4B series test cabinet. The test time is
increased by approx. 250 ms as a result.
Combi-tester G7-1G/-1H/-1T/-1U
Содержание G7-1G
Страница 5: ...5 1 INTRODUCTION free page Combi tester G7 1G 1H 1T 1U...
Страница 38: ...38 4 Operating modes 4 2 Normal Mode Test sequence Combi tester G7 1G 1H 1T 1U...
Страница 39: ...39 4 Operating modes 4 2 1 HV IS test normal mode Combi tester G7 1G 1H 1T 1U...
Страница 40: ...40 4 Operating modes 4 2 2 PE test normal mode Combi tester G7 1G 1H 1T 1U...
Страница 41: ...41 4 Operating modes 4 2 3 Normal mode input step Combi tester G7 1G 1H 1T 1U...
Страница 42: ...42 4 Operating modes 4 2 4 Normal mode output step Combi tester G7 1G 1H 1T 1U...
Страница 43: ...43 4 Operating modes 4 2 4 Normal mode time step Combi tester G7 1G 1H 1T 1U...
Страница 44: ...44 4 Operating modes 4 3 Fast Start Mode test sequence Combi tester G7 1G 1H 1T 1U...
Страница 45: ...45 4 Operating modes 4 3 1 Fast Start Mode PE HV ISO Test Combi tester G7 1G 1H 1T 1U...