Action
Step
NOTICE
If only one membrane strip is removed or if a mem-
brane strip breaks and only part of it is removed, the
connection should not be considered aseptic.
Immediately clamp off this probe sheath assembly on
the short section of silicone tubing between the dispos-
able bag and the female ACD connector.
Remove the anti-actuation ring, which prevents unintentional engagement
before the ACD connectors have been correctly connected.
7
Xcellerex XDUO 2500 Mixer Operating Instructions 29146667 AH
191
6 Preparation
6.4 Prepare and install the probes
6.4.4 Insert probes into the disposable bag