![EG Xcellerex XDUO 2500 Скачать руководство пользователя страница 251](http://html1.mh-extra.com/html/eg/xcellerex-xduo-2500/xcellerex-xduo-2500_operating-instructions-manual_2367561251.webp)
User maintenance schedule
The following table describes maintenance activities that must be scheduled by customer
personnel. All maintenance that is not described in this manual must be performed by
a GE service engineer.
Reference
Performed by
Activity
Frequency
Section 8.5 Cleaning, on
page 257
Operator
Clean the system
Weekly
See the manufacturer's
manual.
Properly trained person-
nel
Change the TCU filter
Annually
Contact your GE represen-
tative.
GE service engineer
Change the I/O Cabinet
vent filter
Verify agitator alignment
If anything abnormal is
found, contact your GE
representative for ser-
vice.
Properly trained person-
nel
Inspect cables and con-
nectors
Annually and after
system relocation
Inspect hinges and han-
dles
Inspect jacking bolts, on
page 259
Properly trained person-
nel
Inspect jacking bolts
If anything abnormal is
found, contact your GE
representative for ser-
vice.
Properly trained person-
nel
Inspect pressure relief
valve
Inspect pump heads
Inspect the I/O Cabinet
Inspect vent and drain
valves
Verify weight monitoring
accuracy, on page 261
Properly trained person-
nel
Verify weight monitoring
functionality
Contact your GE represen-
tative.
GE service engineer
Replace fuses
As needed
Verify agitator speed ac-
curacy
1
1
If the instrument is used less than 12 times a year, the verification can be made annually.
Xcellerex XDUO 2500 Mixer Operating Instructions 29146667 AH
251
8 Maintenance
8.2 Maintenance schedules