Action
Step
For Plus bag assembly: adjust the probe connectors of the disposable bag
by carefully pulling them through the probe window.
11
Adjust impeller position
If the impeller baseplate is not correctly installed, the agitator system is not properly
magnetically coupled and does not function correctly.
To correct the impeller position, follow the instructions below.
NOTICE
Do not disengage the impellers if the liquid level in the disposable
bag is higher than the impeller baseplate.
Action
Step
Disengage the impellers as follows, one by one:
1
a
Pull out the agitator handle completely from under the mixer
vessel as described above.
b
Press the handle slowly downwards. You need to use light force
to overcome the magnetic attraction between the agitator drive
head and the impeller baseplate.
Result: The protective collar surrounding the impeller baseplate is raised.
The impeller baseplate is disconnected from the agitator drive head.
Xcellerex XDUO 2500 Mixer Operating Instructions 29146667 AH
171
6 Preparation
6.3 Prepare the disposable bag