Logged data can be exported to an external computer or to a USB memory stick.
•
Operator actions are recorded automatically.
The following user actions are available to a local user:
•
Adjust temperature calibration offset values.
•
Prime and reverse pumps.
•
Release control to FlexFactory.
•
Start, stop and reset timer.
•
View and adjust alarm settings (available only for Supervisor and Administrator).
•
View and adjust local process tuning parameters (P, I, D).
•
View and adjust local process variables (PV, CV, SP).
•
View Trend screen.
The number of available parameters is dependent on product selection at time of pur-
chase.
FlexFactory component
Two control modes are available in this setup. The following table describes the features
of these control modes.
FlexFactory control
Local control with historization
The process control is centralized.
The process control is local.
The process variables are controlled from
M-Station OIT.
The process variables are controlled us-
ing local control panel.
The local control panel displays the cur-
rent process variables, which cannot be
modified.
The process variables can be modified
using the local control panel.
Full-time trending data is available.
5 min trending data is available.
Historization is available from the
FlexFactory system.
Historization is available via local data
logging and audit trail.
Authorization is needed to make changes to local process values.
Graphical display of the mixer system layout is available.
pH and temperature monitoring and control are available.
Weight and conductivity monitoring are available.
Xcellerex XDUO 2500 Mixer Operating Instructions 29146667 AH
75
3 System description
3.4 XDUO 2500 Mixer control
3.4.1 XDUO 2500 Mixer automation and control