10.4.1
TCU requirements
Value
Parameter
Property
Model B
2
Model A
1
5.2 kW
15 kW
Cooling capacity
General requirements
9 kW
9 kW
Heating capacity
2.8 bar (40 psi g)
2.8 bar (40 psi g)
Fluid outlet pressure
Process requirements
40 L/min
83 L/min
Fluid outlet flow
0°C
0°C
Fluid outlet minimum temperature
65°C
65°C
Fluid outlet maximum temperature
1.5
″
1.5
″
Connector size
4 to 20 mA
4 to 20 mA
Setpoint
Remote control
interface
Required
Required
On/Off emergency stop
1
Supported by XDUO 2500 Mixer.
2
This information is given as an example of a smaller TCU.