9.4
Local control panel
Corrective action
Possible causes
Problem
Connect the power cable.
Power cable is not connected.
The local control panel does not
function.
Contact your GE representative
for replacement cable.
The main power cable is dam-
aged.
Contact your GE representative
for possible replacement.
A fuse has blown.
The home screen does not show
up on the local control panel, al-
though the system is plugged in.
Contact your GE representative
for possible replacement.
A circuit breaker is tripped.
Contact your GE representative
for possible replacement.
The local control panel screen is
non-functional.
A button on local control panel
screen is not working.
Contact your GE representative
for assistance with configuration
adjustment.
The mixer configuration is not
correctly set up.
A specific screen is not accessi-
ble from the local control panel.
Xcellerex XDUO 2500 Mixer Operating Instructions 29146667 AH
269
9 Troubleshooting
9.4 Local control panel