Action
Step
Tap RESET on the local control software screen.
4
The following happens after XDUO 2500 Mixer has been restarted:
Result
System
E Stop Pressed message is no longer shown on the local
control software screen. XDUO 2500 Mixer remains in a safe
state.
Stand-alone mixer
E-Stop Active message is no longer shown on the FlexFactory
Mixer overview screen.
FlexFactory connect-
ed system
•
If a batch was running prior to when the EMERGENCY
STOP was pushed, the system is set in the Held state.
•
If a batch was not running, XDUO 2500 Mixer remains in
a safe state.
If a batch was running prior to emergency stop, the user should verify the status of all
PID control loops and setpoints before restarting the batch after power restoration.
Xcellerex XDUO 2500 Mixer Operating Instructions 29146667 AH
37
2 Safety instructions
2.3 Emergency procedures
2.3.2 Emergency shutdown