Action
Step
Use the navigation keys to make selections in the menu.
6
1
Select Calibrate and press ENTER (2).
2
Select Sensor 2 and press ENTER.
3
Select Cell K.
4
Enter the cell constant value that you found in the accompanying man-
ual.
5
Press ENTER.
6
Press EXIT (3).
Result: The main screen is displayed. The conductivity probe is now ready
to be used.
Disconnect the probe cable from the probe.
7
This procedure is sufficient for most applications. If additional procedures are needed
(changing an offset or resetting the system), see the manufacturer's documentation.
182
Xcellerex XDUO 2500 Mixer Operating Instructions 29146667 AH
6 Preparation
6.4 Prepare and install the probes
6.4.2 Prepare the conductivity probe