4-62
6205s2 BATHYMETRIC AND SIDE SCAN SONAR
0025262_REV_A
3.
Auxiliary data is then combined with the raw sonar data and instantaneous sound velocity (SV)
measurements. They are time-stamped with a common value and then sent back to the 6205s2
Topside Interface Box.
4.
The Discover Bathymetric application takes in the raw sonar data and sends it to the Bathymetric
Processor.
5.
The Bathymetric Processor processes the raw sonar data to produce uncorrected range and angle
data and then transmits this back to the Discover Bathymetric application, where it is packaged
with the auxiliary and dual-frequency side-scan data. At this point, each ping is represented by a
complete data package containing all sonar and auxiliary data.
NOTE:
Once the data is passed back to Discover Bathymetric, data is
displayed on the provided graphical displays to illustrate the bathymetry
and simultaneous dual-frequency side scan data in real-time. The data
may also be recorded as a *.JSF file if the user desires.
6.
The completed data package (range/angle, dual-frequency side scan, and all auxiliary data) is sent
to the third-party software.
7.
The third-party software processes these complete data packets to a native format. The third-
party software also provides graphical displays to show the bathymetry, dual-frequency side scan,
and auxiliary data.
8.
Finally, the third-party software saves native data files that contain every piece of information to
post-process the data and generate final hydrographic products.
Содержание 6205S2
Страница 32: ...2 32 Figure 2 2 6205s2 Sonar Head with Adaptor Flange ICD...
Страница 33: ...2 33 Figure 2 3 6205s2 Side Scan Transducer...
Страница 34: ...2 34 Figure 2 4 6205s2 Side Scan Transducer Details...
Страница 35: ...2 35 Figure 2 5 6205s2 Deck Cable Wiring Diagram...
Страница 36: ...2 36 Figure 2 6 6205s2 GNSS INS Deck Cable Wiring Diagram...
Страница 37: ...2 37 2 6 2 6205s2 Topside Figure 2 7 6205s2 Rack Mounted Topside...
Страница 38: ...2 38 Figure 2 8 6205s2 Rack Mount in Case...
Страница 41: ...3 41 3 1 2 2 Sonar Processor Endcap and Pinout Drawings Figure 3 5 6205S2 Forward End Cap...
Страница 42: ...3 42 6205s2 BATHYMETRIC AND SIDE SCAN SONAR 0025262_REV_A Figure 3 6 6205s2 Aft End Cap...
Страница 46: ...3 46 Figure 3 11 6205s2 Acoustic Center Location X Y Z and Bolt Pattern...
Страница 47: ...3 47 Figure 3 12 6205s2 with Adaptor Flange Acoustic Center Location X Y Z and Bolt Pattern...
Страница 91: ...8 67 Figure 8 22 Checking all COM Ports...