63
Eng
lis
h
Follow these steps to program pauses:
• Access menu level 2, then press the
parameter key to select the special
function .
• The display indicates the symbol for the
special function and the selected pause
durations.
top: abduction/external rotation pause
bottom: adduction/internal rotation
pause (see display image)
The pause duration for abduction/exter-
nal rotation appears in a dotted frame.
• To change the abduction/external rota
-
tion pause, press the ”+“ or ”-“ key. Select
a duration of 10 seconds, for instance.
• Press the
parameter key
again when
done. The dotted frame automatically
moves to the line below, indicating that
the adduction/internal rotation pause
can be adjusted.
The
parameter key
is only used to
toggle between these two functions.
• To modify the value, press the ”+“ (plus)
or ”-“ (minus) key. Select a duration of 5
seconds, for instance.
• Save the programmed values by press
-
ing the
STOP
key, then press the
START
key to begin the treatment.
n
Therapy timer
Default setting is continuous
operation.
A clock symbol in the upper right-hand
corner of the display identifies the
continuous mode of operation.The clock
indicates the elapsed therapy time.
In the continuous mode, the
device must be stopped with the
STOP
key.
However, the therapy timer can be set in
steps of
1 minute to any value between
1 and 300 minutes
.
At the end of the therapy time, the de-
vice automatically stops in the starting
position (maximum adduction position /
halfway between internal and external
rotation).
In this case, a circle replaces the clock
symbol. The circle fills as the therapy
time progresses.
n
Speed
The speed can be adjusted between 1 %
and 100 % in steps of 1 %.
100 % is equivalent to 230°/minute
Default: 100 %
n
New patient
With this function, the CPM device will
move to the home position.
Select the function and press the
START
key:
− The carriage moves to the home
position.
– The programmed treatment param-
eters are deleted.
– All values stored on the chip card are
deleted.
– The device stops halfway between
the abduction/adduction and inter-
nal/external rotation position.
The ”New patient“ function (home posi-
tion) selects the following settings:
– Abduction:
41°
– Adduction:
39°
– Internal rotation:
1°
– External rotation:
-1°
– Pauses:
0