62
LEVEL 1:
n
Abduction
Maximum value:
175 degrees
n
Adduction
Maximum value:
30 degrees
Caution!
Patient hazard —
When using the shoulder restraint, be
sure not to program abduction values
greater than 80°.
n
Internal rotation
Maximum value:
- 90 degrees
n
External rotation
Maximum value:
90 degrees
Note!
• The programmed values and the
values (degrees) measured on the
patient may deviate slightly.
• To ensure a physiological motion, the
carriage simultaneously moves to the
following positions in the synchro-
nized mode:
– Maximum abduction value simul-
taneous with maximum external
rotation value
– Maximum adduction value simul-
taneous with maximum internal
rotation value
• To ensure that the movements you
program are physiological in nature,
we recommend to program the car-
riage in the following order:
1. Maximum adduction value
2. Maximum internal rotation value
3. External rotation
(to relieve the shoulder)
= 0°
4. Maximum abduction value
5. Maximum external rotation value
• In the synchronized mode, after acti
-
vation of
START
the carriage will first
move to the maximum adduction
position and to the position halfway
between internal and external rota-
tion. Then the carriage will stop,
allowing the patient to be positioned.
When the
START key
is pressed again
(start of therapy), the carriage will
first move to the maximum internal
rotation position. Motor A (abduc-
tion/adduction) stops during this pro-
cedure. Once the maximum internal
rotation position has been reached,
both motors (A and B) will simultane-
ously move to the maximum abduc-
tion and external rotation position.
After reaching this position, the cycle
restarts:
maximum adduction and maximum
internal rotation.
• The elevation movement (flexion) is
programmed by means of the abduc-
tion/adduction values (for mechani-
cal adjustment instructions, refer to
section 5.6 ”Examples of use“).
• Horizontal adduction/abduction is
adjusted manually. These are possible
settings:
Horizontal adduction:
120°
Horizontal abduction:
0°
LEVEL 2:
n
Pauses
Pauses occur at each programmed
maximum value.
These are the two pause points:
– Maximum adduction value simulta-
neous with maximum internal rota-
tion value
– Maximum abduction value simulta-
neous with maximum external rota-
tion value
Pauses are adjustable in steps of 1 sec-
ond between 0 and 30 seconds.
Default: no pause
Содержание 80.00.023
Страница 135: ...131 Français ...
Страница 181: ...177 Español ...
Страница 225: ...221 Italiano ...
Страница 270: ...266 ...
Страница 273: ...VII Notes ...