64
– Timer:
continuous
operation
– Speed:
100 %
– Reverse on load -
motor A:
25
– Reverse on load -
motor B:
25
– Motor A:
ON
– Motor B:
ON
– Synchronized mode:
ON
– Total therapy time:
0
– Special functions:
disabled
LEVEL 3:
n
Reverse on load - motor A
(safety circuit)
The motors automatically switch to op-
posite directions of movement when the
patient‘s resistance (load) exceeds the
set value.
Adjustable levels for reverse on load fea-
ture: 1 - 25. At level 1, very low resistance
will cause the device to reverse; at level
25, a high resistance is required to initi-
ate the reversal.
Default: level 25
n
Reverse on load - motor B
(safety circuit)
The motors automatically switch to op-
posite directions of movement when the
patient‘s resistance (load) exceeds the
set value.
Adjustable levels for reverse on load fea-
ture: 1 - 25. At level 1, very low resistance
will cause the device to reverse; at level
25, a high resistance is required to initi-
ate the reversal.
Default: level 25
Caution!
Patient hazard —
The reverse on load feature is a safety
measure to protect the patient in the
event of cramps, spasms, locked joints
and similar situations. The manufac-
turer cannot be held liable for misuse of
this feature.
n
Motor A ON/OFF
To permit fully isolated movements,
the motors can individually be turned
on and off. Motor A effects adduction
and abduction, motor B effects internal
rotation and external rotation.
For an isolated internal / external
rotation movement, set motor A to the
programmed position (abduction/ad-
duction), then turn the motor off.
In normal mode the display indicates the
symbol
”OFF“
during therapy, instead of
the programmed adduction/abduction
angles.
Default: motor A ON
n
Motor B ON/OFF
To permit fully isolated movements,
the motors can individually be turned
on and off. Motor A effects adduction
and abduction, motor B effects internal
rotation and external rotation.
For an isolated adduction/abduction
movement, set motor B to the pro-
grammed position (internal / external
rotation), then turn the motor off.
In normal mode the display indicates the
symbol
”OFF“
during therapy, instead of
the programmed internal/external rota-
tion angles.
Default: motor B ON
Note!
Please note that one motor (A or B)
must always be turned on. Otherwise
the following symbol will be displayed
when you press
START
:
LEVEL 4:
n
Transport setting
With this function, the carriage will
move to a position optimally suited
for packing the CPM device. Select the
function and press the START key. The
carriage moves to the transport posi-
tion. In the top left-hand corner of the
display you will see
(also refer to
section 6.3 Transport).