5.6 Examples of use/
sample protocols
5.6.1 Isolated
adduction/abduction
1. Complete the mechanical settings and
adjust the carriage to the patient as
described in section 4.2.
2. Press the
MENU key
on the program-
ming unit to access level 1 (M1).
3. Press the
parameter key
for
internal
rotation
or external rotation
and
press the
+/- keys
to set the
carriage to the desired rotation position
to be maintained during the isolated
abduction/adduction movements.
4. Next set the desired range of motion
for adduction
and abduction by
first pressing the
parameter key
for
adduction
and set the values with the
+/- keys
.
5. Set the
abduction
value in the same
way
.
Note!
– For an exclusive adduction/abduction
protocol, motor B for the rotation
movement must be turned off as
described in steps 6 to 8 below.
– Having programmed the ranges of
motion, you can program additional
options, such as pause or speed.
6. Press the
MENU key
on the program-
ming unit to access level 3 (M3).
7. Press the Motor B ON/OFF parameter
key
, to activate the parameter.
8. Deactivate motor B by pressing the
parameter key
Motor B ON/OFF
again
or the „–“ (minus) key. The check mark in
the circle next to the symbol must have
disappeared.
9. Save the settings with the
STOP key
, set
the carriage to the starting.
5.6.2 Isolated
internal/external rotation
1. Complete the mechanical settings and
adjust the carriage to the patient as
described in section 4.2.
2. Press the
MENU key
on the program-
ming unit to access level 1 (M1).
3. Press the
parameter key
for
adduction
or
abduction
and press
the
+/- keys
to set the carriage to the
desired position to be maintained during
the isolated rotation movements.
4. Next set the desired range of motion
for internal/external rotation by first
pressing the
parameter key
for
internal
rotation
and set the values with
the
+/-keys
.
5. Set the
external rotation
value in the
same way
.
Note!
– For an exclusive rotation protocol,
motor A for adduction/ abduction
must be turned off as described in
steps 6 to 8 below.
– Having programmed the ranges of
motion, you can program additional
options, such as pause or speed.
6. Press the
MENU key
on the program-
ming unit to access level 3 (M3).
7. Press the
Motor A ON/OFF
,
parameter key to activate the parameter.
8. Deactivate motor A by pressing the
parameter key
Motor A ON/OFF
again
or the
„–“
(minus) key. The check mark in
the circle next to the symbol must have
disappeared.
9. Save the settings with the
STOP key
,
set the carriage to the starting position
with the
START key
and initiate therapy
by pressing the
START key
again.