54
4. Adjustment of elbow angle
5. Adjustment of forearm length
6. Adjustment of headrest
When the optional patient re-
straint is used¹
7. Adjustment of headrest and shoul-
der restraint¹
− It is not permitted to change the
adjustment of 1 through 7 while a
patient is sitting in the CPM device.
− Movements must
not cause any pain
or irritation
.
− Patients must be
fully conscious
while being instructed in the use of
the CPM device and during therapy.
− The choice of the
therapy param-
eters
to program and of the
therapy
protocols
to use is restricted to the
responsible physician or therapist.
It is the physician‘s or therapist‘s de-
cision whether or not to use the CPM
device on a specific patient.
− The patient must be familiar with the
functions of the
ARTROMOT®-S3
programming unit
and the unit must
be
within easy reach
of the patient,
allowing him or her to stop therapy, if
needed.
Patients unable to operate
the programming unit
, e.g. paralytic
patients, must never be left unat-
tended during therapy.
− Write the patient‘s name on the
patient chip card
. The card should
only be used for this patient. If the
patient chip card is used for another
patient, be sure to
delete the previ-
ous patient‘s data
from the card first
(see sections 4.1 and 5.3, paragraph
”New Patient“). Use
original chip
cards only
.
− All accessories used with the
ARTROMOT®-S3
device must first be
approved by DJO Global.
− Modifications to the medical device
described in this document without
the manufacturer’s written consent is
prohibited.
− Ensure that the medical device is
securely set up on solid ground during
use.
− Do not allow
parts of the body or
any objects
(such as blankets, cush-
ions or cables) to get caught
in the
moving parts
of the CPM device.
Warning!
− Extreme caution should be taken
when in use around small children and
babies! Sufficient distance to the device
is mandatory for their safety!
− Never leave the device unattended
when it is switched on! Switch the
device off and disconnect the power
cord from the wall outlet!
− After use, store the device in a safe place!
Ensure device stability also during storage!
Do not rest or lean on or push the
assembled device because it might tip
over.
Warning!
Shock hazard —
Strictly observe the following warnings.
Failure to do so endangers the lives of
the patient, the user and other persons
involved.
− Before use allow the
ARTROMOT®-S3
to reach room temperature. If the
device has been transported at
tem-
peratures below 0 °C (32 °F)
, leave it
to dry at room temperature for about
2 hours, until any condensation has
disappeared.
− The
ARTROMOT®-S3
must only be
operated in dry rooms.
− When disconnecting the device from
the power line, remove the plug from
the wall outlet first, before disconnect-
ing the cable from the device.
− When connecting the device to other
equipment or when creating a medical
system, check that the sum of leakage
currents will not cause any hazard.
Please contact DJO Global, if you have
questions in this matter.
− Do not use multiple portable socket
outlets (MPSO) to connect the device
to the power line.
ARTROMOT®-S3
must be connected to a
properly in-
stalled wall outlet with a non-fused
earthed wire
. Before connecting the
power cord, it must be completely
unrolled and placed such that it will
not get caught in the moving parts of
the device.