72
5.6.3 Isolated elevation
(flexion)
1. First adjust an angle of 90° for external
rotation and deactivate motor B as
follows:
a. Press the MENU key on the program-
ming unit to access level 1 (M1).
b. Press the
parameter key
for
external
rotation
and move the car-
riage to the rotation position of 90°,
using the
+ / - keys
.
Note!
– For an exclusive elevation protocol,
motor B for the rotation movement
must be turned off as described in
steps c to e below.
– Having programmed the ranges of
motion, you can program additional
options, such as pause or speed.
c. Press the
MENU key
on the program-
ming unit to access level 3 (M3).
d. Press the
Motor B ON/OFF
parameter key to activate the
parameter.
e. Deactivate motor B by pressing the
parameter key Motor B ON/OFF again
or the „–“ (minus) key. The check mark
in the circle next to the symbol must
have disappeared.
f. Save the entered values by press the
STOP key
.
2. Complete the mechanical settings and
adjust the carriage to the patient as
follows:
a. Adjust an angle of 90° for horizontal
adduction/abduction
†
, using index-
ing knob (12).
b. Secure and hold the motion element
at the holding tube for length adjust-
ment and open the locking screw
(14) for height adjustment
Œ
. Select
a height at which the pivot points of
motor A and the shoulder are at the
same level.
c. Now open clamping lever (18) for ad-
justment of the upper arm length
Å
,
locking screw (20) for adjustment of
the elbow angle
è
and locking screw
(28) for adjustment of the forearm
rest inclination and set the motion
element to a position that is comfort-
able to the patient and suitable for
the therapy session . When done,
tighten all screws and levers.
d. Open clamping lever (26), adjust the
forearm length
^
to the patient and
close the clamping lever.
3. In the next step, program the elevation
range of motion as follows:
a. Press the
MENU key
on the program-
ming unit to access level 1 (M1).
b. Next set the desired range of mo-
tion for adduction and abduction by
first pressing the
parameter key
for
adduction
and set the values
with the +/- keys.
c. Set the
abduction
value in the
same way.
Note!
Having programmed the ranges of
motion, you can program additional op-
tions, such as pause or speed.
4. Save the settings with the
STOP key
,
set the carriage to the starting position
with the
START
key
and initiate therapy
by pressing the
START key
again.