INSTRUCTIONS FOR USE DA 270, DA 370, DA 280, DA 380
UM_EN_DA270_DA280_DA370_DA380_m2015_2019-04_ver1.9 61 / 67
9.
pour the remaining diluted solution (ca 1l) from the Oro Cup into the spittoon bowl and rinse away with a small
amount of water
10. it is recommended to carry out the disinfection of the suction and waste system with the Orotol Plus agent at least
once a day (preferably at the end of each working day)
11.
at the beginning of the next working day, suck 1l of clean cold water through the hose of the big and small aspirator
10.8
Cleaning, disinfection and decontamination of other parts of the dental unit
•
Clean the unit’s external surfaces including the glass surface of the keyboard
and the leatherette of the chair
upholstery with a wet cloth
•
Manufacturer recommends to use
Incidin™ Foam
- spray (HENKEL - ECOLAB)
•
Cleaning should be carried out regularly or when the surface of the product has become contaminated with biological
material incidentally.
•
It is forbidden to clean the upholstered parts of the unit with agents that have detrimental effect upon
the structure of the leatherette, such as acetone, trichloro, perchloro, alcohol, abrasive cleaning
agents, polishing agents
•
Do not use phenol- and aldehyde-based cleaning agents to clean other parts of the unit;
they can permanently damage the structure of varnish and plastic surfaces
Clean the following parts at least once a day
(according to the version):
•
sieve of the separation block located in the spittoon block
•
sieve at the input to the amalgam trap
•
sieve of the saliva ejector
•
filter of the small and big aspirator
•
sieve in the spittoon bowl
Instruments and handpieces
Cleaning, disinfection and sterilization of the instruments and their handpieces should be
carried out according to the instructions of their manufacturer, furnished with the instrument.
When cleaning the floor (PVC floor) with a cleaning agent, it is forbidden to place the foot controller on the
wet floor
The manufacturer shall not be taken responsible for damages arisen due to the use of other disinfecting
and cleaning agents than those recommended.