INSTRUCTIONS FOR USE DA 270, DA 370, DA 280, DA 380
UM_EN_DA270_DA280_DA370_DA380_m2015_2019-04_ver1.9 55 / 67
10
CLEANING, DISINFECTION AND DECONTAMINATION
10.1
Disinfection of the cooling waterlines of instruments
It is recommended to use Alpron
or Sanosil S003 or Dentosept P agent in 1% concentration with distilled water. The
solutions are poured into the reservoir for distilled water and can be used permanently. The agents in 1% concentration
are harmless to the health of the patient. With regular use the cooling system is maintained in cleanliness and it is not
necessary to use other disinfecting agents. Contact your dealer for details about the possibilities of the purchase and use
of the said products.
In the event that water from the central distribution is used for the cooling of instruments, the disinfection of the internal
distributions of the instruments is made as follows:
1. Fill the reservoir of distilled water with 1% solution of Alpron or Sanosil S003 or Dentosept P with distilled water.
2. Turn the triple-
throw switch to the position „DISTIL“
3. Rinse the water path of an instrument for 30s, other instruments that use cooling water are rinsed for 10s.
4. Turn the triple-throw switch i
n the spittoon block to the position „CENTRAL“
The manufacturer recommends to carry out the above disinfection at least once a day, preferably at the end of the
working day.
10.2
Semi-automatic disinfection of the cooling waterlines of instruments
(optional)
Control of the hygiene from the touch screen of the dentist table.
To get access to HYGIENE mode, press the button HYGIENE in SETTINGS screen.
The hygiene screen is displayed:
The only active button is: