INSTRUCTIONS FOR USE DA 270, DA 370, DA 280, DA 380
UM_EN_DA270_DA280_DA370_DA380_m2015_2019-04_ver1.9 56 / 67
After the button has been pressed, the following is displayed in the lower part of the screen:
Fit the bottle with the disinfecting solution and then confirm the activity by pressing the button with the message. After the
button has been pressed, a new message is displayed:
Insert the funnel (Figure a) into the container.
Fit the container prepared in described way over the spittoon bowl (Figure b).
Take out all instruments for the disinfection of the cooling waterlines.
The instruments should be taken off the hoses and the hoses should be inserted into the beds in the container on the
spittoon bowl.
Syringe is controlled by autonomous valve. You have to perform filling of its hose manually:
Put the tip of the syringe to the instruments holder, press the left button and hold until the disinfection solution
flows out from the tip. Put the syringe back to its holder.
Until now it is possible to cancel the process of the waterlines disinfection by pressing button ESC.
By pressing the button START the process of waterlines filling with disinfecting solution starts.
REPLACE THE BOTTLE WITH DETERGENT
INSERT HOSES INTO HOLDER