INSTRUCTIONS FOR USE DA 270, DA 370, DA 280, DA 380
UM_EN_DA270_DA280_DA370_DA380_m2015_2019-04_ver1.9 26 / 67
Button for general enabling of alarm after the lubrication time of the instrument has been reached.
Press the button to enable / disable the indication of the lubrication time of the instrument being reached.
The condition is indicated by means of the symbol
- turned ON /
- turned OFF. The button
enables/ disables the indication of the lubrication for all instruments simultaneously. Which instrument
should be indicated is set individually for each instrument in the service menu.
Button for turnig the automatic bowl flushing on/off at bowl returning to basic position
.
Press the button to enable/disable automatic bowl flushing at its returning to the rest position.
The status is indicated by symbol
- turned ON /
- turned OFF.
Button for turnig the automatic bowl flushing on/off at bowl moving from basic position
.
Press the button to enable / disable automatic bowl flushing at its move from the rest position. The status is indicated by
symbol
- turned ON /
- turned OFF.
Turn ON/OFF the dental light automatically after chair returns back from Getting-on position
When this option is selected (ON), then after the chair returns back from getting-on position,
the light automatically returns to its previous setting i.e. it will glow in the brilliance of what was
set before. If this option is not selected (OFF), then the light will not turn on, after the chair returns back from getting on
position. That means that dentist may turn it on manually. (e.g. when the next patient is coming).
Going to the previous page of the user menu.
Press the button to display the previous page of the user menu.
Returning to the main screen
Press the button
„ESC“ to return to the main screen.
Menu SERVICE / MANUFACTURER
This option serves for service and production setting of the parameters. Particulars can be found in the Instructions for
Service.
Menu TEST
This option serves to test some functions of the dental unit.