INSTRUCTIONS FOR USE DA 270, DA 370, DA 280, DA 380
UM_EN_DA270_DA280_DA370_DA380_m2015_2019-04_ver1.9 57 / 67
If all the hoses have been inserted into the holder, confirm activity by pressing Start button.
The process is indicated by the message:
The time in seconds to the end of the process is displayed in the lower part of the button.
After the cooling of the instruments has been filled up with the disinfecting solution, the solution starts to act.
The status is indicated by the message:
The process can be ended at any time by pressing the button INTERRUPTION or by turning the dental unit off. The duration
of the treatment with the solution is left to the operator’s discretion and should take whole night or at least 60 minutes.
After pressing the button INTERRUPTION or after turning the unit on again (after the night), the following is displayed:
Fit the bottle with distilled water and confirm the activity by pressing the button with the message.
After the button has been pressed, a new message is displayed:
The time in seconds to the end of the rinsing process is displayed in the lower part of the button.
Due to the fact that syringe is controlled by autonomous valve, its rinsing is necessary to be performed
manually.
Confirmation by pressing the button will end the process of the disinfection of the instruments and the hoses can be
returned into the dentist table.
After rinsing has been finished, new message is displayed:
START
FILLING
29
ACTIVITY ACTIVITY
INTERRUPT
ACTIVITY
ACTIVITY
P
REPLACE THE BOTTLE WITH WATER
FLUSHING
01: 29
END