INSTRUCTIONS FOR USE DA 270, DA 370, DA 280, DA 380
UM_EN_DA270_DA280_DA370_DA380_m2015_2019-04_ver1.9 51 / 67
Intraoral camera
The camera does not serve to make diagnosis but only for better visualization in the dental treatment.
The camera comprises of:
•
holder
•
connection
– connector
•
the instrument itself (camera)
The product must be protected against water. Keep it dry.
8.4
Dental operating light
Dental operating light can be controlled by means of the three-position selector switch on the dental operating light, while
in the middle position the light is turned off.
If the unit is fitted with electronic control of illumination, it is possible to control it from the contzrol panel and the assistant
arm by means of a button- see the description of the buttons.
For the use of the dental operating light, please read the manual for the light in question.
8.5
Cart (only for DA 280, DA 380
–
CART)
The cart with the control panel is connected to the spittoon block by means of a supply harness in a protective hose. The
hose protects the harness against mechanical strain, dust, humidity, against abrasion and at the same time it forms
a supplementary insulation. The outer sheath of the hose is made of soft PVC.
The protective hose contains a built-in inner coil made of hard PVC, qhuich makes it non-yielding in the event an adult
person treads upon it. However, the manufacturer recommends not to stand on the hose or place any object on it. The
protective hose must not be bent to a radius less than 70 mm. To clean the hose, follow the instructions in the chapter
„Cleaning, disinfection and decontamination of other parts of the dental unit“
.
Button to control the lift blocking (CART)
In the CART version, the working height of the control panel can be changed within the range of 200 mm. Grip the control
panel (preferablyx while standing behind it) from the sides with both hands, while pressing (and holding) with a finger the
button located on the bottom side of the panel, thus unblocking the mechanism of height adjustment. Re-position the panel
gripped in the described way to the desired working height. Block the panel in the set height by releasing the mechanism
button.
When moving the cart, ensure that the connection cable of the foot controller or any other object does not get
in touch with the castors of the cart