3. OPERATION
Page 73
© 2009 DH Instruments, a Fluke Company
For best results, wait 5 minutes with the PPCH vented before controlling
pressure when changing to or from dynamic control mode. This is to allow time for
the PPCH thermal pressure control unit to respond and stabilize.
Control mode setting is range specific. A change in control mode made while in
one PPCH range does NOT affect the control mode setting in other ranges.
The current control mode is indicated by two characters in the middle of the
bottom line of the MAIN RUN screen (<D> for standard dynamic, <d> for hi volume
dynamic, <S> for static, <M> for monotonic, <R> for ramp, with <C> appended if
custom control settings are in use). (See Section 3.1.1)
3.4.6.3
TURNING-OFF CUSTOM CONTROL PARAMETERS
To return to default pressure control parameters for the active range and control
mode, select the control mode using
[SETUP]
,
<6control>
,
<2mode>
(see
Section 3.4.6). Selecting the mode automatically activates its default control
parameters. This is a convenient way to put the PPCH into a known condition
after trying custom control settings.
3.4.7
<7DRV> (DRIVERS)
PURPOSE
To control the output signals of PPCH’s 8 channel, 12 V external drivers.
PRINCIPLE
PPCH external drivers are available to drive peripheral equipment in a PPCH system, for
example, solenoid valves. The driver electrical connections are available from a rear panel
connector.
See
Section
7.1 for driver specifications and connector pin-outs.
OPERATION
To access the driver control function press
[SETUP]
,
<7drv>
. The display is:
External drivers: 1
2 3 4 5 6 7 8
Pressing the keypad numerical key driver number turns that driver ON and OFF with either a
momentary or a toggled response. An active driver is indicated by
<*>
immediately following
the driver number.
Pressing
[ENT]
while in the external drivers menu causes a menu to appear that allows
selection of whether the driver actuation by selecting the driver number will be
<1momentary>
or
<2toggle>
.
Содержание PPCH
Страница 10: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 106: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 96 N NO OT TE ES S...
Страница 144: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 182: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 172 N NO OT TE ES S...
Страница 190: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 180 N NO OT TE ES S...