3. OPERATION
Page 77
© 2009 DH Instruments, a Fluke Company
When AutoZ is run for an external Q-RPT in an RPM4, the PPCH causes the RPM4 to run its
own AutoZero routine. The RPM4 Q-RPT is then AutoZeroed as if it had been AutoZeroed
independently of the PPCH.
To access the PPCH AutoZ function menu press
[SPECIAL]
,
<1AutoZ>
. The display is:
1.
Active Q-RPT designator.
2.
Indication of active measurement mode (
<abs>
for absolute,
<gage>
for gauge).
3.
Indication of whether AutoZ is currently ON or OFF for this
Q-RPT and measurement mode.
1off 2view IH
3edit 4run abs ON
Select
<1off>
(or
<1on>
) to change the AutoZ status for the current Q-RPT and
measurement mode from ON to OFF or vice versa.
AutoZ ON is indicated by a <z> in the MAIN RUN screen, top line, fourth character from the
right. When AutoZ is OFF, the character is blank.
Select
<2view>
to view the current value of
P
offset
for the active Q-RPT and measurement
mode.
P
offset
should be zero for Axxx Q-RPTs when the PPCH is new or has just been calibrated.
P
offset
should be roughly equal to atmospheric pressure for a PPCH Q-RPT in gauge mode.
Select
<3edit>
to edit the value of
P
offset
for the active Q-RPT and measurement mode
(see Section 3.5.1.1).
The value of P
offset
is always displayed and entered in Pascal (Pa). Normally, in day to day
operation, the value of P
offset
is changed by running AutoZ (see Section 3.5.1.2), not direct editing.
Select
<4run>
to run the AutoZ routine which determines and activates
P
offset
relative to
P
std,0
(see Section 3.5.1.2).
<4run>
is only applicable to absolute measurement mode. AutoZ is
run automatically without operator intervention in gauge measurement mode whenever
PPCH is in the VENT condition (see Section 3.2.4).
3.5.1.1
EDIT AUTOZ
The edit AutoZ function should be used with great caution as entering
inappropriate values and turning ON AutoZ may result in out of tolerance
measurements. In normal day to day operation, the value of the AutoZ offset, P
offset
,
should be changed using the run AutoZ function (seen Section 3.5.1.2) in absolute
measurement mode or automatically in gauge modes.
Before editing P
offset
, see Section 3.5.1,
PRINCIPLE.
Содержание PPCH
Страница 10: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 106: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 96 N NO OT TE ES S...
Страница 144: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 182: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 172 N NO OT TE ES S...
Страница 190: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 180 N NO OT TE ES S...