3. OPERATION
Page 91
© 2009 DH Instruments, a Fluke Company
3.5.7
<7INTERNAL>
PURPOSE
To view, set, adjust, and maintain various aspects of PPCH’s internal operation.
OPERATION
To access the internal selections press
[SPECIAL]
,
<7internal>
. The display is:
1config 2baro
3pump
The INTERNAL menu choices include:
<1config>
To run automated routines that readjust PPCH internal pressure control
coefficients and determine the test volume to which the PPCH is connected
(see Section
3.5.7.1).
<2baro>
View the real time output of the on-board barometer (see
Section
3.5.7.2).
<3pump>
Run an automated pump priming routine, view the status of the pump supply
reservoir level switch, view the reading of the pump output pressure
transducer (see Section
3.5.7.3).
3.5.7.1
<1CONFIG>
PURPOSE
To run automated routines that determine control valve opening coefficients and
the test volume to which PPCH is connected.
PRINCIPLE AND OPERATION
These functions are considered part of PPCH maintenance and are therefore
covered in the maintenance section of this manual (see Sections 5.7, 5.8).
3.5.7.2
<2BARO>
PURPOSE
To view the value of atmospheric pressure as measured by the PPCH on-board
barometer.
PRINCIPLE
PPCH is equipped with an on-board barometer. The atmospheric pressure
measurements made by the on-board barometer are used for dynamic
compensation of atmospheric pressure when using an Axxx Q-RPT to make gauge
pressure measurements (see Sections 3.3.3,
PRINCIPLE and 3.2.4). The
barometer, if calibrated, may also be used as a reference for AutoZ of Q-RPTs in
absolute measurement mode (see Section 3.5.1).
The on-board barometer is a low accuracy sensor used only for measuring
small changes in atmospheric pressure over short periods of time
(see Section 3.2.4) PPCH measurement uncertainty does not depend on the
measurement uncertainty of the on-board barometer unless the barometer is used
as a reference for AutoZ of Q-RPTs in absolute measurement mode.
Содержание PPCH
Страница 10: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 106: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 96 N NO OT TE ES S...
Страница 144: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 182: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 172 N NO OT TE ES S...
Страница 190: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 180 N NO OT TE ES S...