PPCH™ OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 170
The utility sensor is for pressure indication, safety and housekeeping functions.
It is a temperature compensated micro-machined silicon sensor module or a
strain gauge transducer.
5.14.2.3
TEST VENT VALVE
Open/close need valve. The valve is normally closed by the force of two
Belleville washer. It is opened by an opposing force applied by pneumatic
pressure on a bladder.
5.14.2.4
LO Q-RPT SHUTOFF VALVE
Same as the test vent valve (see Section 5.14.2.3).
5.14.2.5
LO Q-RPT
PPCH may be equipped with a Lo Q-RPT. The Lo Q-RPT provides a lower
measurement range than the Hi Q-RPT or utility sensor. See Section 5.14.2.2
for a description of Q-RPTs and Section 1.2.2.1 for complete Q-RPT
specifications.
5.14.2.6
LO Q-RPT VENT VALVE
Same as the test vent valve (see Section 5.14.2.3).
5.14.2.7
PUMP ACCUMULATOR
High pressure stainless steel volume of approximately 40 cc.
5.14.2.8
FRONT PANEL
Sheet metal with inlaid overlay panel on which a membrane switch keypad and
electro-luminescent display are mounted.
5.14.2.9
ELECTRICAL CABLE TRACK
Articulated track to organize various cables connecting between the PPCH’s
lower mechanical unit and upper electrical unit. The track is articulated so that
the mechanical unit can slide out.
5.14.2.10
HYDROPNEUMATIC PUMP
The pump is a pneumatically powered, gate valve controlled, piston pump similar to
a double-acting pneumatic cylinder. The pump pistons have a nominal ratio of
400:1 (PPCH-200, PPCH-100) or 200:1 (PPCH-140, PPCH-70) between the
pneumatic piston and hydraulic plunger depending on PPCH model. Drive
pressure to the pump is controlled by the drive pressure regulator
(see Section 5.14.2.16). The pump operates continuously until the pneumatic
drive pressure on the pneumatic piston is in equilibrium with the oil pressure on
the oil plunger.
5.14.2.11
INLET CONTROL VALVE
Highly progressive needle valve. The valve is operated by a stepping motor and
drive belt.
Содержание PPCH
Страница 10: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 106: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 96 N NO OT TE ES S...
Страница 144: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 182: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 172 N NO OT TE ES S...
Страница 190: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 180 N NO OT TE ES S...