PPCH™ OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 4
1.2.2.2
UTILITY SENSOR
Base PPCH pressure controllers include a utility sensor. If a Hi Q-RPT
(see Section 3.2.5) is installed, there is no utility sensor.
Utility sensors are used for pressure control, system monitoring and safety
functions. They are intended for indication only, not to provide reference
pressure measurement. Q-RPTs (see Section 1.2.2.1) are used for reference
measurement.
Warm Up Time
None required
Range
PPCH-70M:
70 MPa (10 000 psi) absolute and gauge
PPCH-100M: 100 MPa (15 000 psi) absolute and gauge
PPCH-140M: 140 MPa (15 000 psi) absolute and gauge
PPCH-200M: 200 MPa (30 000 psi) absolute and gauge
Resolution
Precision
0.001 % of span
0.1 % of span
1.2.2.3
ON-BOARD BAROMETER
The on-board barometer is used only to measure changes in atmospheric
pressure to provide dynamic compensation of the Q-RPT’s atmospheric pressure
offset when operating in gauge measurement mode. If calibrated,
it can also be used as an AutoZ reference for AutoZeroing in absolute
measurement mode.
Warm Up Time
None required
Range
70 to 110 kPa (10 to 16 psi)
Resolution
0.001 % of span
Precision
Predicted Stability
0.1 % of span
0.05 kPa/yr (0.008 psi/yr)
1.2.3
PRESSURE CONTROL SPECIFICATIONS
ALL MODELS
Control Precision (Standard Dynamic Mode)
to ± 0.003 % of Q-RPT span
Default Dynamic Control Hold Limit
± 0.01 % of AutoRanged range
Lowest Controllable Pressure
(absolute,
gauge modes)
1 MPa (150 psi)
Typical Control Ready Time (Dynamic Mode)
60 to 120 seconds.
Slew Rate (0 to FS)
60 seconds
TEST Volume
0 to 100 cc, 50 cc optimum (operates in larger
volumes but pressure stabilizing time increases)
Содержание PPCH
Страница 10: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 106: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 96 N NO OT TE ES S...
Страница 144: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 182: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 172 N NO OT TE ES S...
Страница 190: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 180 N NO OT TE ES S...