PPCH™ OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 110
HS%
Purpose
“Static” or “dynamic” control modes: Read or set the automated pressure control hold limit as
% of range.
“Monotonic” control mode: Read or set the “target limit” as a % of range.
Command
Query
“HS%
limit
”
“HS%?”
Classic “HS%=
limit
“
“HS%”
Arguments
Limit:
The limit in % span of the active range.
Remarks
The “hold” or “target” limit can be read and set as a percent of the span. The “hold” limit
applies to “static” and “dynamic” control modes only. These limits are part of the ready criteria.
The “target” limit applied to “monotonic” control mode only. If this program message is used
to set the hold or target limit, the PPCH will then use CUSTOM control settings
Example
(enhanced)
Cmd sent:
Reply:
“HS%.01”
“0.0100 %”
(No reply if GPIB-488)
Example
(enhanced)
Cmd sent:
Reply:
“HS%? 0.01”
“0.0100 %”
Example
(classic)
Sent:
Reply:
“HS%= 0.01”
“0.0100 %”
Errors ERR#
6
The
‘limit’
argument was invalid.
See Also
“HS”, “MODE”
3.2.2.2, 3.2.2.4
ID
Purpose
Read or set the user defined instrument identification label.
Command
Query
“ID
string
”
“ID”
Classic
“ID=
string
”
“ID”
Arguments String:
An alphanumeric string up to 12 characters long.
Remarks
The user defined ID label can be used to allow the user to “tag” the PPCH with a unique
identifier. This ID is stored in non-volatile memory and cannot be erased by a power failure,
system fault or reset
Example
(enhanced)
Cmd sent:
Reply
“ID PPCK #A01”
“PPCK #A01”
(No reply if GPIB-488)
Example
(enhanced)
Cmd sent:
Reply
“ID? PPCH #A01”
“PPCH #A01”
Example
(classic)
Sent:
Reply:
“ID=PPCH #A01”
“PPCH #A01”
Errors
ERR# 6
The
‘string’
argument was longer than 12 char.
See Also
3.5.5.4
IF
Purpose
Increase the pressure quickly.
Command “IF
n
”
Classic “IF=
n
”
Arguments ‘0’
‘1’
Stops the pressure generation.
Starts an ascending generation at a fast fixed rate.
Remarks
The up control valve is used to establish a fixed rate of 2% FS/sec. The valve is closed as the
pressure approaches the target. Care must be used, as the pressure will not stop increasing
until the valve is closed, or the upper limit is exceeded
Example
(enhanced)
Cmd sent:
Reply:
“IF 1”
“1”
(no reply if IEEE-488)
Example
(enhanced)
Cmd sent:
Reply:
“IF? 1”
“1”
Example
(classic)
Sent:
Reply:
“IF=1”
“IF=1”
Errors ERR#
6:
ERR# 55:
The
n
argument is a ‘0’ or a ‘1’.
Hardware malfunction, fluid reservoir is low.
See Also
“DF” , “DS”, “IS”, “IP”, “DP”,
3.2.1
Содержание PPCH
Страница 10: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 106: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 96 N NO OT TE ES S...
Страница 144: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 182: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 172 N NO OT TE ES S...
Страница 190: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 180 N NO OT TE ES S...