3. OPERATION
Page 65
© 2009 DH Instruments, a Fluke Company
Press
[ENT]
to save the range or
[ESC]
to return to the MAIN RUN screen without
saving the range.
Default (DF) Q-RPT ranges cannot be saved. They are permanent.
Duplicate AutoRange ranges cannot be saved. A duplicate AutoRange range is a
range that has the same measurement mode and full scale and uses the same Q-RPT
as a range that has already been saved.
3.4.1.2
DELETING A
UTO
R
ANGE
RANGES
OPERATION
To delete a single saved AutoRange range, press
[SETUP]
,
<1range>
,
<2delete>
. If there are any saved ranges available to delete, the display is:
1.
Indication that the cursor control keys may be
used to toggle through the saved ranges to
select the one to be deleted.
2.
Designator and position of the Q-RPT used by
the range to be deleted.
3.
Pressure unit of measure of the range to be
deleted.
4.
Full scale pressure of the range to be deleted
in gauge mode and absolute mode if available.
< and > A200M X1H
MPa 170g/170a
Use the cursor control keys to toggle through the ranges available to be deleted.
When the desired range to be deleted is displayed, press
[ENT]
. If you are sure you
want to delete this range, press
[ENT]
again in response to the
<Delete?>
query.
Press
[ESC]
to return to the
<AutoRange:>
screen without deleting a range.
To delete all of the saved AutoRange ranges, press
[SETUP]
,
<1range>
,
<3delete all>
. If you are sure you want to delete all the ranges, respond
<1yes>
to the
<Delete all saved ranges?>
query. If not, respond
<2no>
.
3.4.2
<2RES> (RESOLUTION)
PURPOSE
To set the resolution with which measured pressures and other pressure indications and
settings are displayed.
PRINCIPLE
The resolution with which PPCH displays pressure values can be adjusted.
This feature can be used to reduce the resolution when lower precision measurements are
being made and excess resolution might confuse or distract the operator.
The resolution setting determines the number of digits with which pressure is displayed.
The resolution is calculated based on the full scale of the range and then rounded to the
furthest digit to the right. For example, resolution of 0.001 % on a range of 150 MPa is
150 X 0.001 % = 0.0015 which is rounded down to 0.001 MPa.
Default resolution is 10 ppm of active range span. Resolution is set automatically by
AutoRange (see Section 3.3.4).
Содержание PPCH
Страница 10: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 106: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 96 N NO OT TE ES S...
Страница 144: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 182: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 172 N NO OT TE ES S...
Страница 190: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 180 N NO OT TE ES S...