PPCH™ OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 16
1. PPCH
2.
2.75 in. (70 mm) DH 500 nipple
3. DH500
tee
4.
12 in. (305 mm) DH500 nipple
5.
DH500 tee (RPM4) or union (PG7000)
6.
6 in. (152 mm) DH500 nipple
7.
DH500 union (RPM4) or tee (PG7000)
8.
5 in. (127 mm) DH500 x 1/8 in. tube
weld nipple (slightly flexible)
9a. RPM4 or RPM3
9b. PG7000
Figure 6.
Connecting PPCH to an RPM4 or PG7302 using the standard
PPCH interconnection accessories
2.3.9.3
CONNECTING TO A PG7302 PISTON GAUGE
ALWAYS use external tubing and fittings rated for pressure equal to or
greater than the maximum pressure PPCH will be used to generate.
PPCH is frequently used as the automated pressure generation and control
component in a PG7302 oil operated piston gauge system. To do so, it must be
connected hydraulically and by RS-232 to the PG7302.
The PG7302 puts the PPCH into remote mode and manages its operation.
To connect a PPCH to a PG7302 to be used as an automated pressure
generation and control component in the PG7302 system, proceed as follows:
Using tubing of appropriate pressure rating and a tee fitting, connect the
PG7302
TEST
port to the PPCH
TEST
port (see Figure 6 for instructions on
making this connection using the standard interconnections accessories
supplied with the PPCH). The third leg of the tee is for the connection to the
device or system under test.
Using a standard pin-to-pin DB-9M to DB-9F RS232 cable, connect COM1 of
the PPCH to COM3 of the PG7302.
See the PG7000 Operation and Maintenance Manual for information on
operating the PG7302 with an automated pressure controlling device.
For PPCH to be identified by a PG7302 platform, the PPCH-COM1 port settings
must be the same as the PG7302 COM3 settings (baud rate, parity, data bits, stop
bits). If the PG7000 is not able to establish communications with PPCH, check that
the PPCH COM1 port settings are the same as the PG7302 COM3 settings. If the
COM port settings are correct, check that the correct communications cable is being
used (standard pin-to-pin DB-9M to DB-9F RS232) and is connected to the correct
communications ports.
See the PG7302 Operation and Maintenance Manual for additional information on
PG7302 RS232 communications and COM port settings.
Содержание PPCH
Страница 10: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 106: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 96 N NO OT TE ES S...
Страница 144: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 182: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 172 N NO OT TE ES S...
Страница 190: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 180 N NO OT TE ES S...