3. OPERATION
Page 29
© 2009 DH Instruments, a Fluke Company
3.2.2.2
DYNAMIC CONTROL
Dynamic control
mode is designed to set the pressure to the target value and
control continuously to keep pressure within the hold limit and as close to the
target value as possible (see Figure 10).
The advantage of this control mode is that the final pressure achieved is the
same as the target value. The maximum value of the control error is equal to the
hold limit. The average value of the control error is generally much smaller than
the hold limit.
During dynamic pressure control, the hold limit is active. If the pressure goes
outside of the hold limit, a
Not Ready
condition occurs. See Table 10 for default
hold limit values and Section 3.4.6 to customize the hold limit.
There are two types of dynamic control.
Standard
dynamic control is the default
type of dynamic control used in most situations.
Hi volume
dynamic control is
also available for situations in which standard dynamic control cannot achieve a
Ready
condition in a reasonable amount of time and/or cannot maintain the
Ready
condition once achieved.
Hi volume
dynamic control may be necessary
when the test volume connected to the PPCH is greater than 100 cc. The
difference between
standard
and
hi volume
dynamic pressure control is in the
use of the thermal pressure controller (TPCU). In
standard
dynamic control,
control transitions from the control valves to the TPCU before
Ready
occurs and
the target pressure is reached. In
hi volume
dynamic control, control is by the
valves only unless they valves are unable to maintain the
Ready
condition.
Figure 10.
Dynamic pressure control operation
PPCH includes a
Ready/Not Ready
feature that indicates when current pressure
control requirements have been met (see Section 3.2.3).
With dynamic pressure control active, a
Ready
condition occurs whenever:
•
The current measured pressure is inside the hold limit.
Figure 11 illustrates
Ready/Not Ready
operation in dynamic control mode.
In dynamic control mode, normal procedure is to assume that when a
Ready
condition occurs (pressure inside the control hold limit), the measured pressure
equals the target pressure. For this reason, when the pressure is Ready in dynamic
control mode, the measured pressure display is equal to the target pressure.
Содержание PPCH
Страница 10: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 106: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 96 N NO OT TE ES S...
Страница 144: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 182: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 172 N NO OT TE ES S...
Страница 190: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 180 N NO OT TE ES S...