PPCH™ OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 72
1.
Active Q-RPT designator.
2.
View/edit field of monotonic ramp rate limit in
active pressure unit of measure per second.
Shows active rate limit.
3.
Active control mode (
<M>
for monotonic).
Rate limit: IH
0.050 MPa/s M
Edit the target ramp rate if desired and press
[ENT]
to return to the main run
screen with the edited values active.
See Section 3.4.6, Table 10 for default control limits.
Ramp Control Mode
When ramp control mode is active, the
[SETUP]
,
<6control>
,
<1limits>
display is:
1.
Active Q-RPT designator.
2.
View/edit field of pressure ramp rate hold limit
value in active pressure unit of measure per
second.
3.
Active control mode (
<R>
for ramp).
Rate hold limit: X1H
0.100 MPa/s R
Edit the ramp rate hold limit value if desired and press
[ENT]
to return to the
main run screen with the edited values active.
See Section 3.4.6, Table 10 for default control limits.
3.4.6.2
<2MODE>
PURPOSE
To select the active automated pressure control mode.
PRINCIPLE
See Section 3.4.6,
PRINCIPLE.
OPERATION
The control mode selection display is:
1dynamic 2static X1H
3monotonic 4ramp
The cursor is on the active control mode. Making a control mode selection
activates the control mode, sets default control parameters for that control mode
(see Section 3.4.6, Table 10) and returns to the main run screen. To set custom
control parameters, press
[SETUP]
,
<6control>
,
<1limits>
from the MAIN RUN
screen (see Section 3.4.6.1).
Dynamic control mode selection, standard or hi volume
When
<1dynamic>
is selected, the type of dynamic control (standard or hi
volume) must be specified (see Section 3.2.2.2).
After selecting
<1dynamic>
, display is:
Dynamic control X1H
1standard 2hi volume
Select the desired dynamic control mode type. The default is
<1standard>
.
Making a control mode selection activates the control mode, sets default control
parameters for that control mode.
Содержание PPCH
Страница 10: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 106: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 96 N NO OT TE ES S...
Страница 144: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 182: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 172 N NO OT TE ES S...
Страница 190: ...PPCH OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 180 N NO OT TE ES S...